Centrifuge Tubes: Sizes, Materials and How to Choose the Right One

A centrifuge tube is a plastic tube made to survive being spun, with a maximum force printed on its datasheet. Most bench work runs on four sizes: 0.2 mL PCR tubes, 1.5 and 2.0 mL microcentrifuge tubes, and 15 and 50 mL conical tubes. Choose the size by your volume, the material by what goes inside and how hot or cold it gets, and the tube by its RCF rating, which must sit well above the force your protocol asks for.
The other half of the decision is the machine. A tube is only useful if your rotor, or an adapter in it, holds it snugly, and if you know the force in × g rather than just the rpm on the display. Below are the choices one at a time, then a simple cell pellet spin worked through.
The sizes you will meet most
| Size | Usual shape | Typical jobs |
|---|---|---|
| 0.2 mL PCR tubes, single or strips of 8 | thin walls, domed or flat caps | PCR and qPCR set-up, small reactions |
| 1.5 mL microcentrifuge tubes | conical bottom, snap cap | aliquots, small dilutions, DNA and protein preps |
| 2.0 mL microcentrifuge tubes | often round bottom, snap or screw cap | volumes just too big for 1.5 mL, mixing |
| 15 mL conical tubes | conical bottom, screw cap, printed scale | pelleting cells, splitting samples |
| 50 mL conical tubes | conical, sometimes with a skirt | media, buffers, pooled cell harvests |
PCR tubes and strips
Thin walls carry heat quickly between the cycler block and a reaction of 10 to 50 µL. Domed caps are common for ordinary PCR, while qPCR usually calls for flat, optically clear caps, since most instruments read the signal through them. A few seconds in a mini centrifuge brings droplets down off the walls; these tubes are not meant for long, hard spins.
Microcentrifuge tubes
The 1.5 mL tube is the most common plastic item on a molecular biology bench. Its conical tip collects a small pellet where you can find it. Polypropylene microtubes from the large makers are often rated to 20,000 × g or more: Corning lists 20,000 × g for one of its 1.5 mL tubes, and Eppendorf gives up to 30,000 × g for some of its locking tubes.
Conical tubes
The 15 mL conical is the standard vessel for spinning cells out of one flask; the 50 mL size takes media, buffers and pooled harvests. The printed scale is fine for reading roughly how much is there. For an accurate volume, measure with a pipette; our guide to using a micropipette covers the small volumes.
Polypropylene or polystyrene
| Polypropylene (PP) | Polystyrene (PS) | |
|---|---|---|
| Clarity | translucent; "high clarity" grades are clearer | glass clear |
| Solvents | excellent with dilute acids, bases and alcohols; not recommended for aromatic or halogenated hydrocarbons | limited solvent resistance |
| Heat | maximum use about 130 °C | maximum use about 90 °C; melts in an autoclave |
| Cold | brittle at about 0 °C | brittle at room temperature |
| Example rating | 12,000 × g (Corning 15 mL PP conical) | 1,800 × g (Corning 15 mL PS conical) |
The figures come from Thermo Fisher's plastics selection pages and Corning's catalog. Polypropylene is the default for anything spun hard, frozen or exposed to more than a buffer. Polystyrene earns its place when you need to see clearly through the wall, at low force, with aqueous samples. Chemical resistance also depends on temperature, time and centrifugal stress, so check a resistance chart before spinning solvents.
Caps and bottoms
Snap caps are quick and suit microtubes you open often. Screw caps are the usual choice for samples that travel, go into the freezer or contain anything hazardous. Never cap a tube with foil; university safety guides such as UC San Diego's warn that it can come off or tear during a run.
Conical bottoms gather a pellet into a point, and round bottoms are easier to resuspend in. A skirted 50 mL tube stands on the bench without a rack, but check that your adapters take the skirt.
Sterile, DNase/RNase-free and other claims
"Sterile" means the packaged tubes were sterilized, usually by gamma irradiation or ethylene oxide (EO). "Non-pyrogenic" means endotoxin tested below a stated limit; Corning tests its Falcon tubes to less than 0.1 EU/mL. "DNase/RNase-free" means lots tested below the detection limit of a particular assay. These are separate claims: a sterile tube is not automatically RNase-free, or the reverse. Match the claim to the job: sterile and non-pyrogenic for cells going back into culture, nuclease-free for RNA.
RCF ratings, and why rpm alone is not enough
A centrifuge displays rpm, how fast the rotor turns. Your sample feels relative centrifugal force (RCF, written × g), which depends on rpm and the distance from the center of rotation:
RCF = 1.118 × 10^-5 × r × rpm²
with r in centimeters to the bottom of the tube. Force grows with the square of the speed and in proportion to the radius. A mini centrifuge at 7,000 rpm with a rotor radius near 5 cm gives about 2,680 × g. The same 7,000 rpm on a 15 cm benchtop rotor gives about 8,200 × g, three times as much. That is why an rpm copied from someone else's protocol can crush cells or leave a loose pellet. The rpm to rcf calculator converts in either direction once you know your rotor's radius.
Two ratings apply to every run, the tube's and the rotor's. Use the lower one, and remember that the rotor manual may lower the allowed speed again for some adapters or dense solutions. A rating is a ceiling, not a target.
Balancing the rotor
- Load tubes in opposite pairs of matched mass, not just matched volume. A dense solution and a light one can look level and still be out of balance.
- Weigh pairs when it matters. Some university guides ask for opposing tubes within 0.1 g.
- With an odd number of samples, add a balance tube of water.
- Opposite each other, use the same tube type, cap and adapter.
- Check tubes for cracks before each run, and stay with the machine until it reaches speed. If it shakes or rattles, stop it.
Freezing and autoclaving cautions
Water expands about 9 % as it freezes, so a full, tightly capped tube can split. Leave headspace and freeze tubes upright in a rack, as Sarstedt's freezing guidance advises. Plastics are weaker below 0 °C, so let frozen tubes thaw before you spin or vortex them.
Polypropylene conicals are commonly kept at minus 80 °C; Corning rates its 15 mL PP tube for long-term storage there. Ordinary centrifuge tubes are not cryovials, though. University safety offices, Yale's among them, warn that liquid nitrogen can seep into vials kept in the liquid phase and expand violently as they warm. Freezing cells for storage is its own process, outlined on our stem cell technology page.
As a material, polypropylene tolerates a standard autoclave cycle (Thermo Fisher gives 121 °C at 15 psi for 20 minutes for empty containers, caps fully loosened). Polystyrene melts. Whether a particular tube and cap can be autoclaved is a product claim, so read that tube's datasheet. A pre-sterilized tube does not need autoclaving before use.
Labeling that lasts
Label before you fill and before anything goes cold, because labels stick poorly to frosty plastic. Write on the side, where the label stays with the tube, and add a short code on the cap for finding it in a rack. Alcohol wipes smear some marker inks, so use markers or labels sold for lab and freezer use.
Matching the tube to the rotor
Fixed-angle rotors hold tubes at a set angle, and the pellet forms on the outer wall near the bottom. Swing-out rotors let buckets swing horizontal, so the pellet forms in the tip, which is why cell work usually uses them. A 15 mL tube in a 50 mL bucket needs an adapter shaped for a conical tip, and it should sit fully down without wobbling. Microcentrifuge rotors take 1.5 and 2.0 mL tubes, with adapters for smaller ones. Mini centrifuges take microtubes and PCR strips only.
Worked example: tubes for a simple cell pellet spin
You have harvested one T75 flask and have about 9 mL of cell suspension. The protocol says 300 × g for 5 minutes. Your benchtop centrifuge has a swing-out rotor, and its manual gives r max as 16 cm.
| Decision | Choice | Reason |
|---|---|---|
| Size | 15 mL conical | 9 mL fits with headspace |
| Material and claims | sterile, non-pyrogenic polypropylene | the cells go back into culture |
| Rating check | rated 12,000 × g | far above 300 × g |
| Speed | 1,300 rpm | from the formula |
| Balance | a 15 mL tube with 9 mL of water | weighed against the sample |
The speed: rpm = √(300 ÷ (1.118 × 10^-5 × 16)), about 1,295. Setting 1,300 rpm gives roughly 302 × g. Put both tubes in matching adapters in opposite buckets, weigh the pair and adjust the water with a pipette if they differ.
After the spin, remove the supernatant, resuspend the pellet and take 10 µL for a count with the hemocytometer calculator. A mini centrifuge would not have done this job: it cannot hold a 15 mL tube, and its top speed of about 2,680 × g is far above the few hundred × g cells usually get.
Common mistakes
- Setting rpm from a protocol written for a different rotor.
- Spinning a polystyrene tube at polypropylene speeds.
- Balancing by eye with different tube types on opposite sides.
- Overfilling, so the cap and threads get wet.
- Spinning or vortexing tubes straight out of the freezer.
- Treating "sterile" and "RNase-free" as the same claim.
Stocking a small bench
A basic culture or molecular bench gets by with three tube sizes and one small spinner. For 1.5 mL work, a pack of 500 EO-sterilized microcentrifuge tubes lasts a long time. For cells, 15 mL conical tubes in polypropylene with screw caps, EO sterile, non-pyrogenic and DNase/RNase free, come 50 to a pack, with a scale from 2 mL in 1 mL steps and a writing spot. Their 50 mL companions, 25 to a pack, share the specification with a scale in 2.5 mL steps. The seller rates both to 12,000 × g, plenty for cell pellets and routine spins. The listing is unclear on autoclaving, so do not plan on it.
A mini centrifuge rounds out the bench. One common model runs up to about 7,000 rpm, about 2,680 × g, comes with two rotors and takes microtubes and PCR strips. It cannot take 15 or 50 mL tubes. These picks were chosen from specifications, manuals and published reviews.
FAQ
What is the difference between rpm and RCF on a centrifuge?
Rpm is how fast the rotor turns. RCF, written × g, is the force on the sample, and it depends on both rpm and rotor radius.
What size centrifuge tube should I use for cells?
For one flask's worth of cells, a sterile 15 mL polypropylene conical is the usual choice. Larger or pooled harvests go into 50 mL conicals.
Can a mini centrifuge spin 15 or 50 mL tubes?
No. Mini centrifuges hold microtubes and PCR strips. Conical tubes need a benchtop centrifuge with buckets or adapters for them.
Can I freeze samples in centrifuge tubes?
Polypropylene tubes are widely used at minus 20 and minus 80 °C. Leave headspace, freeze upright and thaw before spinning. For liquid nitrogen, use cryovials made for it.